Model: | ABBOTT CD-3700 | Usage: | Medical Clinical |
---|---|---|---|
Origin: | China | Expiry: | 24 Months |
Recycling: | Disposable | Shipment: | Normal Temperature By Sea |
Specimen: | Blood | ||
High Light: | human biochemistry reagents,chemically pure reagents |
Cell Counter Reagents for ABBOTT CD3700 CD3500 Hematology Analyzer Diluent Lyse Sheath Rinse Compatible
Intended Use of Abboot Reagents To be used for blood cell counting, sizing, and WBC differentiation, hemoglobin determination on ABBOTT CD3700 CD3500 CD3000 Hematology Analyzers.
Specification of Abboot Reagents
CBC Reagents compatible on Abbott CD3700 CD3500 CD3000 | ||
Cat. No | Description | Pack Size |
DWX-50807 | DILUENT-CD5 | 20L |
DWX-50808 | LYSE-HGB/WIC | 3.8L/5L |
DWX-50809 | SHEATH-CD37 | 10L |
DWX-50810 | RINSE-CD | 20L |
DWX-50811 | CLEAN-CD | 50ml×2 |
PRECAUTIONS
Re-calibration is recommended when changing the origin of the reagents.
It is not recommended using different brands of solutions together on one unit of analyzer.
Do not ingest. Avoid contact with skin and eyes.
Skin contact: Flush with plenty of water.
Eyes contact: Flush with plenty of water. Obtain immediate medical attention.
This kit is for professionals and in vitro diagnostic use only.
Never put distilled water on the analyzer (except clean for shipping menu), otherwise the pneumatic unit can be damaged seriously.
STORAGE
Store the product at 2-35℃ in a dust free place with good ventilation. Temperature outside the range may damage the product.
The open vial life of the product is 60 days. Unused reagents should be discarded after 60 days. Do not mix residual with a new carton.