Introduction
The HCV Rapid Test (Serum/Plasma) is a rapid visual immunoassay for the qualitative, presumptive detection of antibodies to HCV in human serum or plasma specimens. This kit is intended for use as an aid in the diagnosis of HCV infection.
Principle of Assay
The HCV Rapid Test detects antibodies to HCV through visual interpretation of color development in the internal strip. Protein A is immobilized on the test region of the membrane. During testing, the specimen reacts with recombinant HCV antigen conjugated to colored particles and precoated onto the sample pad of the test. The mixture then migrates through the membrane by capillary action and interacts with reagents on the membrane. If there are sufficient HCV antibodies in the specimen, a colored band will form at the test region of the membrane. The presence of this colored band indicates a positive result, while its absence indicates a negative result. The appearance of a colored band at the control region serves as a procedural control, indicating that the proper volume of specimen has been added and membrane wicking has occurred.
Details
Production Name |
HCV |
Specimen Used | Serum/Plasma |
Storage | Room Tempreture(2~30℃) |
Format | Cassette |
Shelf Life | 24 Months |
Contents
Operation Direction
Bring tests, specimens, and/or controls to room temperature (15-30°C) before use.
1. Remove the test from its sealed pouch, and place it on a clean, level surface. Label the device with patient or control identification. For best results, the assay should be performed within one hour.
2. Using the provided disposable pipette, fill the narrow portion of the pipette with the serum or plasma specimen, transfer 2 drops of serum/plasma to the specimen well (S), then start the timer.
Avoid trapping air bubbles in the specimen well (S), and do not add any solution to the result area.
As the test begins to work, color will migrate across the membrane.
3. Wait for the colored band(s) to appear. The result should be read at 10 minutes. Do not interpret the result after
4. Interpretation of results:
POSITIVE: The presence of two lines as control line (C) and test line (T) within the result window indicates a positive result.
NEGATIVE: The presence of only control line (C) within the result window indicates a negative result.
INVALID: If the control line (C) is not visible within the result window after performing the test, the result is considered invalid.
Features
Accurate • Rapid • Simple
Quality Control
• Internal procedural controls are included in the test. A colored band appearing in the control region (C) is considered an internal positive procedural control, confirming sufficient specimen volume and correct procedural technique.
• External controls are not supplied with this kit. It is recommended that positive and negative controls be tested as a good laboratory practice to confirm the test procedure and to verify proper test performance.